BIBO
The bag in/bag out filtration system (BIBO) with double in situ serves for high-efficiency filtration of supply and exhaust air in bio-safety laboratories. The filter replacement with the help of the bag in/bag out system prevents the personnel from suffering any harm caused by pathogenic factors.
![�豸12.jpg](https://static.wixstatic.com/media/8a6e04_1e1f3e7869a743a4baaa637d66a1d08e~mv2.jpg/v1/fill/w_940,h_525,al_c,q_85,usm_0.66_1.00_0.01,enc_avif,quality_auto/%EF%BF%BD%E8%B1%B812.jpg)
COMPONENTS
Cabinet
Filter unit
Bio-safety airtight valve
Aerosol injection detection port
Pressure test system
Scanning detection unit
Bag in and bag out unit
Aerosol mixing degree sampling unit
Fumigation and disinfection unit
Differential pressure – signal feedback
SPECIFICATIONS
Air Volume Cabinet Size
2000 m³ - 16000 m³ according to the air volume
​
The BIBO with a double in situ is available in a variety of models according to the air volume. The air volume ranges from 2000 m³ to 16000 m³. There are 8 types of standard models with the following air volume: 2000 m³, 4000 m³, 6000 m³, 8000 m³, 10000 m³, 12000 m³, 14000 m³, 16000 m³.
​
Customized models can be manufactured.
​
FEATURES
Made of 304 or 316L stainless steel, it can resist routine laboratory disinfection and corrosion.
The product has a firm structure and resists a pressure of 2500 Pa.
The airtightness meets the requirements of the ASME AG-1 2003 ‘Code on Nuclear Air and Gas Treatment’ with zero leakage as well as the EN 12266-2:2002 Class A standard.
The system is equipped with a high-efficiency filter with a filtration efficiency of 99.97% for particles with a particle diameter of 0.3 μm.
The system is equipped with an in situ detection function which allows the monitoring of the filtering effect without disassembling the equipment.
The aerosol sampling port can exactly locate the leak and has an anti-winding function.
The high-efficiency filter can be replaced safely with the bag in/bag out system. The replacement is easy to handle with the tow hook at the filter installation place. It can be completed by one or two persons.
The standard disinfection interface serves to disinfect the equipment in situ.
The filter resistance-detection function allows a display on-site as well as a remote signal to remind users to replace the filter in time.
​
Conditions of use
ambient temperature 0°C-40°C
relative humidity not more than 85% (no condensation)
environmental pressure 86kPa-106kPa
​